HA 5100 (Automated 5-part Differential Hematology Analyzer)

HA 5100 is first innovative analyzer which combines the optical method of BASO(BASO-O)and impedance method of BASO(BASO-I)together,it can make much more stable and reliable mesurementof sample with pathologic BASO,which helps reduce a lot of the failure analysis by using the analyzer.

Technical Specifications


Feature Value
Principles
  • Multi angle Laser Scatter
  • Flow Cytometry method
  • 3D Scattergram analysis
  • Impedance method for RBC and PLT Counting
  • Impedance method for RBC and PLT Counting
  • Cyanide free reagent for HGB test
Parameters 25 Reportable Parameters:
WBC,RBC,HGB,HCT.MCV,MCH,RDW-SD,RDW-CV,PLT,MPV,PCT,
PDW,P-LCR,P-LCC,NEU%,LYM%,MON%,EOS%,BAS,NEU#,LYM#,
MON#,EOS#,BAS#
25 Reportable Parameters:
ALY%,ALY#,IG%,IG#
1 3D Scattergram or 2 2D Scattergrams
3 Histograms(WBC/BASO,RBC,PLT)
Display 14 inch Touch Screen,resolution 1366*768
Sample Volume CBC+5-Diff model: ≤20ul
CBC model: ≤10ul
Data Memory UP to 100,000 results(including histigram,scattergram,patient information)
Interface 1 LAN port,4 USB ports
Communication Bi-direction LIS supports LIS supports HL7 protocol
Internal RFID reader(optional)
Printout External printer by USB or WIFI connection(Optional)
Throughput 60 tests/hour
Dimension L*W*H = 480*375*517(mm)
Weight:33kg
Power requirment a.c.100-240V,50/60Hz
Working Environment
  • Temp:10-30oC
  • Humidity:20% - 85%
  • Air Pressure: 70∼1065Pa
  • Working latitude:≤3500m
Parameter Linearity Range Carry Over CV
WBC 1.0-99.9X109/L ≤0.5% ≤2.0%
RBC 0.3-7x1012/L ≤0.5% ≤1.5%
HGB 20-240g/L ≤0.5% ≤1.5%
PLT 20-999X109/L ≤1.0% ≤4.0%


PDF Downloads


Brochure Bio-Rad World wide Report
Cost Per Test In Euro CE Certificate
Cross Contamination Report by Rallis Agrochemicals User Manual
Product Comparison by Elitech, France
ecs2000